Buying a medical device importer or distributor: the MDEL and the records that come with it
Section 44(1) of the Medical Devices Regulations says no person shall import or sell a medical device unless the person holds an establishment licence, subject to the exceptions in section 44(2). A buyer of an importer or distributor checks that the licence is held and what records and procedures come with it.
A business that imports or sells a medical device in Canada must hold an establishment licence under the Medical Devices Regulations, subject to the exceptions in section 44(2). This guide covers who needs a medical device establishment licence (MDEL), how to check the seller’s licence, what changes when a business changes hands, the distribution and complaint records, fees and the annual review, and what follows a suspension or cancellation. It covers the federal rules only, and provincial rules may apply to a particular business as well, so ask your lawyer.
Which businesses need an MDEL, and which do not
Section 44(1) of the Medical Devices Regulations says no person shall import or sell a medical device unless the person holds an establishment licence. Section 44(2) says that requirement does not apply to the importation or sale of a medical device by a retailer, a health care facility, the manufacturer of a Class II, III or IV device, or the manufacturer of a Class I device if the manufacturer imports or distributes solely through a person who holds an establishment licence. Section 1 defines a health care facility as a facility that provides diagnostic or therapeutic services to patients, and the definition includes a group of such facilities that report to one common management that has responsibility for the activities carried out in those facilities.
Health Canada says an MDEL is issued for the activities of importing or distributing (selling) medical devices for human use in Canada. Its MDEL FAQ answers a question about a single yearly sale this way: any person who manufactures (Class I), imports or sells a medical device for human use in Canada requires an MDEL. The same FAQ lists those who are exempt, including retailers and healthcare facilities. It also says the requirement to hold an MDEL does not depend on the quantity sold per year.
The FAQ describes a distributor as a person, other than a manufacturer, an importer or a retailer, who sells a medical device in Canada for the purpose of resale or use, other than for personal use. Health Canada’s guidance overview defines a supplier as any person, other than the manufacturer, who distributes (sells) a medical device to an MDEL holder for the purpose of import or sale in Canada. Under section 44(3), a person who imports a medical device shall ensure that the person from whom they import it holds an establishment licence. Section 44(4) says subsection (3) does not apply where a person imports a Class I device from its manufacturer if the person importing it holds an establishment licence, or imports a Class II, III or IV device from its manufacturer. For the short version, see who needs a medical device establishment licence in Canada.
Checking the seller’s licence before you sign
Health Canada’s MDEL page links to a Medical Devices Establishment Licence Live Listing. Ask Health Canada or your lawyer what that listing shows about the seller’s licence. Then ask the seller for the licence as issued and a copy of the application behind it, and compare what they say with what the business actually sells.
An MDEL application includes a statement of whether the applicant is the importer or distributor (or both) and whether it manufactures Class I devices that it distributes. The application also states the class (Class I, II, III or IV) of each device the applicant will import or distribute. It includes an attestation signed by a senior official for the documented procedures that must be in place depending on the activities conducted at the establishment. The attestation needed depends on whether the applicant is an importer, distributor or manufacturer. The application must also include the site addresses of each building where the documented procedures that were attested to are in place.
Where a business lists a site that is not the same legal entity, Health Canada’s guidance says the licence holder must ensure that the listed sites have the applicable procedures in place and that inspectors are able to verify compliance without impediment. Ask who runs each listed site and whether the seller’s arrangement with that site is part of the deal.
What happens to the licence when the business changes hands
The sources read for this guide do not say whether an MDEL moves to a buyer, so ask Health Canada and your lawyer how your deal affects the licence before you sign. What they do cover is the notice a licence holder gives when information changes.
Section 48 says that if, following the issuance of an establishment licence, there is a change to the name and address of the establishment (paragraph 45(a)) or to the name, title and contact information of the representative to contact (paragraph 45(b)), the holder of the licence shall submit the new information to the Minister within 15 days of the change. Health Canada’s guidance says licence holders must notify it within 15 calendar days of a change in contact information. It lists among those changes a change in the name or address of the establishment and a change in the information of the representative associated with the application, including name, title, email and phone number.
Health Canada says you must inform it of other changes affecting the information on your MDEL. For example, if the list of manufacturers, activities or class of devices change, you must submit the information through your annual licence review (ALR) application. Health Canada’s application guidance gives examples of changes as changing the activities or classes of medical devices, sites and the list of manufacturers or suppliers. To submit those changes before the next ALR, an amendment application using FRM-0292 is available only from April 1 to mid-November. Health Canada says there are no fees associated with making changes, notifications or amendments to your MDEL.
A licence holder must inform Health Canada if it chooses to cancel the MDEL. Before it submits the cancellation application using FRM-0292, all activities under the MDEL must have stopped. Only the contact person or senior official for the MDEL may submit a cancellation request to Health Canada. Health Canada says it may inspect an establishment whose MDEL was cancelled to verify that all licensable activities have ceased. Ask the seller who is named as the contact person and senior official today.
The documented procedures you inherit
For distributors of all classes of medical devices, Health Canada’s guidance says the attestation should state that the establishment has written operating procedures, and the list on that page includes maintaining distribution records and handling complaints and recalls. For importers of all classes, the attestation should state that the establishment has written operating procedures, and the list on that page includes handling complaints and recalls, maintaining distribution records and reporting incidents. For importers of Class II, III or IV devices, the attestation should also state that the establishment has written operating procedures for submitting information for any serious risk of injury to human health and for properly handling, storing, delivering, installing, correcting and servicing medical devices (where applicable). For distributors of Class II, III or IV devices, the attestation should also state that the establishment has written operating procedures for properly handling, storing, delivering, installing, correcting and servicing medical devices (where applicable).
Health Canada says it may, at any time, ask for written procedures. It also says that if the procedures attested to in the MDEL application are not available during an inspection, it could consider this a “false attestation”, which may lead to an MDEL suspension. Ask the seller for the written procedures the senior official attested to, and have your lawyer confirm they exist and are in use.
Distribution records and how long they must be kept
Section 52(1) says the manufacturer, importer and distributor of a medical device shall each maintain a distribution record in respect of each device. Subsection 52(2) says subsection (1) does not apply to a retailer, or to a health care facility in respect of a medical device that is distributed for use within that facility. Health Canada’s guidance says you must also have a documented procedure for how you maintain your distribution records, as attested to in the MDEL application.
Section 53 says the distribution record shall contain sufficient information to permit the complete and rapid recall of the medical device. Health Canada’s guidance puts it as enough information to quickly withdraw any medical device completely from the market. Section 55 says the manufacturer, importer and distributor shall retain the distribution record for the longer of the projected useful life of the device and two years after the date the device is shipped. Section 56 says distribution records shall be maintained in a manner that will allow their timely retrieval.
The guidance says procedures should specify how long you retain your distribution records and how they will be maintained, and it points to sections 52 to 56 of the Regulations for the rules, including minimum retention periods and maintaining records to allow quick retrieval. Ask the seller how far back the records go and where they are kept, and whether you can retrieve a device’s record by looking it up.
Complaint files, recalls and incident reports
Health Canada’s guidance says manufacturers, importers and distributors must also maintain records of reported problems concerning the performance or safety of all medical devices they have sold, and that these records must include all actions taken to respond to the problems. Section 57(1) says each must maintain records of reported problems relating to the performance characteristics or safety of the device, including any consumer complaints, received after the device was first sold in Canada, and of all actions taken in response. Section 57(2) exempts a retailer, and a health care facility in respect of a device distributed for use within that facility. Section 58 says each shall establish and implement documented procedures that enable an effective and timely investigation of those problems.
Health Canada’s guidance says manufacturers, importers and distributors must have documented procedures for recalls. It says that when they recall a medical device, the manufacturer and importer must each send a report to Health Canada on or before the recall, and that distributors should comply with section 45(g) and conduct a recall when required. As soon as possible after completing a recall, both the manufacturer and importer must each report to Health Canada the results of the recall and the actions taken to prevent a recurrence of the problem.
Section 63.2 says a manufacturer or importer who decides to recall a device without being ordered to do so by the Minister shall provide the Minister with specified information, in writing, within 24 hours after making the decision. Section 63.1 says sections 63.2, 64 and 65 do not apply to a manufacturer or importer unless the device is likely to cause injury to the health of a patient, user or other person, or could cause serious injury to the health of a patient, user or other person. Health Canada’s guidance notes that section 59 sets incident reporting requirements for manufacturers as well as importers.
Health Canada recommends that manufacturers and importers develop internal written procedures for reporting medical device shortages and discontinuations, and a suggested element is a procedure for regularly checking the list of medical devices subject to notification of shortages so that all mandatory shortages are reported. Ask for the complaint file, any recall or incident reports, and the written procedures behind them.
Records and procedures to ask the seller for
| Item | What the Regulations or Health Canada say |
|---|---|
| Distribution record | The manufacturer, importer and distributor of a device each maintain one for each device; subsection 52(2) exempts a retailer and a health care facility in respect of a device distributed for use within that facility |
| Retention of that record | The longer of the projected useful life of the device and two years after the date the device is shipped (s. 55) |
| Complaint and problem records | Records of reported problems, including consumer complaints, received after the device was first sold in Canada, and of the actions taken in response (s. 57(1)); subsection 57(2) exempts a retailer and a health care facility in respect of a device distributed for use within that facility |
| Complaint procedure | Documented procedures that enable an effective and timely investigation of those problems (s. 58) |
| Recall procedures | Manufacturers, importers and distributors must have documented procedures for recalls, according to Health Canada’s guidance |
Fees and the annual licence review
Health Canada says fees are charged for the initial MDEL, subsequent annual licences and the reinstatement of an MDEL, and that the fees are applied equally to foreign and domestic establishments. The same page says some establishments are exempt from MDEL fees. The MDEL fee increases annually by an amount equivalent to the Consumer Price Index (CPI) from the previous year, beginning on April 1, 2021. Each fall, Health Canada says it will publish a Notice of Intent in the Canada Gazette that specifies the fees to be implemented the following April 1.
All active MDEL holders must submit an application for Annual Licence Review (ALR) prior to April 1 of each year. The ALR must be submitted even if there are no changes to the content of the MDEL application. The licence continues to be valid as long as the ALR application is submitted before April 1 of each year and the associated fee is paid. If the ALR application is not submitted before April 1, Health Canada says it will cancel the MDEL starting on April 1, at which point the holder is no longer authorized to conduct licensable activities. If your deal closes near the annual review deadline, ask the seller to show that the ALR has been filed and the fee paid. The shorter answers on renewing a medical device establishment licence and on the licence fee cover these points on their own.
If the licence is suspended or cancelled
If an MDEL is suspended or cancelled, Health Canada’s guidance says the holder must immediately stop importing or selling medical devices. Health Canada says it may suspend an MDEL when it has reasonable grounds to believe that the holder has contravened the Regulations or any provisions of the Food and Drugs Act related to medical devices, that the holder made a false or misleading statement when applying for the licence, or that not suspending the MDEL would constitute a risk to the health or safety of patients, users or other persons.
A holder may apply to reinstate a suspended licence to resume selling or importing, but must correct the problems that caused the suspension, and Health Canada says it will inspect the establishment to verify the corrections before it decides. The Minister or delegated authority must cancel an MDEL when the licence has been suspended for a period of more than 12 months, or when no annual licence review application is submitted before April 1 of each year. Health Canada’s authority to reinstate an MDEL is limited to suspended MDELs, and if an MDEL is cancelled and the establishment wishes to resume activities, it must apply for a new MDEL and meet the requirements set out in section 45 of the Regulations, as applicable. If cancelled following a suspension, the holder must first correct the problems that caused the suspension and cancellation, and submit documentation with the application to demonstrate that this has been done.
Questions to ask before you buy
- Does the seller hold an MDEL today, and which activities and device classes does it cover?
- Has the annual licence review been filed on time every year, and has the fee been paid?
- Who is the named contact person and senior official, and will they stay through the sale?
- Where are the written procedures the senior official attested to, and are they in use?
- Can the seller show distribution records for each device and say where they are kept?
- What complaints, recalls and incident reports are on file, and what was done about each?
- Has Health Canada inspected the business, suspended the licence or raised a problem?
- How does my deal affect the licence, and who confirms that with Health Canada?
Questions people ask
Short answers; each links to the full page.
- Who needs a medical device establishment licence in Canada?
- Under the federal Medical Devices Regulations, no person imports or sells a medical device unless the person holds an establishment licence, subject to the exceptions in section 44(2), including retailers and health care facilities. Health Canada’s guidance says importers and distributors need an active MDEL for all classes of device, subject to the listed exemptions. Full answer
- Is there a fee for a medical device establishment licence?
- Health Canada says fees are charged for the initial MDEL, subsequent annual licences and reinstatement, and that some are exempt. Its FAQ says there is a fee for the annual licence review application and no fee for amending or notifying an MDEL. Full answer
- How do I renew a medical device establishment licence?
- Health Canada says an MDEL does not expire annually and continues to be valid as long as the holder submits its annual licence review application before April 1 each year and pays the associated fee. If the application is not submitted before April 1, it says it will cancel the MDEL starting on April 1. Full answer
- How do I check whether a medical device is licensed in Canada?
- Health Canada keeps a database of all licensed Class II, III and IV medical devices offered for sale in Canada. It can be searched by company name or ID, licence name or number, and device name or identifier. Class I devices do not appear in it. Full answer
Related reading
14 pages on the same subject, grouped by topic.
Legal
- DefinitionMedical device establishment licence (MDEL)
- DefinitionMedical device licence
- GuidePatient records, licensing and regulatory approval in a practice sale
- ChecklistLicence and permit transfer checklist
- ChecklistPermits and licences renewal checklist
- DefinitionLicence re-application
- DefinitionAsset sale
- DefinitionShare sale
Due diligence
Sources
Every requirement and figure referenced in this guide traces to a primary source. Links were last confirmed on the dates shown.
- 01Health CanadaGovernmentMedical Device Establishment Licences
- 02Government of Canada (Department of Justice)GovernmentMedical Devices Regulations, SOR/98-282
- 03Health CanadaRegulatorGuidance on medical device establishment licensing (GUI-0016): Overview
- 04Health CanadaRegulatorGuidance on medical device establishment licensing (GUI-0016): How to apply for an MDEL
- 05Health CanadaRegulatorGuidance on medical device establishment licensing (GUI-0016): How to maintain your MDEL
- 06
- 07
- 08Health CanadaRegulatorFrequently asked questions: Medical device establishment licensing and fees
- 09Health CanadaRegulatorFees for Medical Devices